How to Select Closed-Loop Grab Sampling System

SENSOR Sampling System Accessories

SENSOR Sampling Systems designs and manufactures closed-loop grab sampling systems. With SENSOR sampling systems, customers eliminate potential hazards by extracting representative grab samples safely, while also managing quality and yield. Special sample collection technology lets these systems capture hazardous emissions during sample collection and keep those emissions from exposing personnel or being released to the environment.

Sample coolers, valves, emission filters, and enclosures are accessories to these systems. Each one should match the process rather than come from a generic catalog.

Mismatched accessories cause problems. For example, if a cooler does not cool properly, the station cannot be used. Viton™ is not compatible with H2S, so a stream with H2S needs a different elastomer, or the Viton™ fails early. A bottle that does not fit the shroud lets the vent needles strike the cap or bend. Each of these could trace back to missing process or application data. Specify accessories from process data instead of part numbers.

 

Looking to solve one of these common issues:

  • Sample running too hot to handle safely or damaging delicate instruments: see Sample Coolers
  • Needle bending, cap not sealing, or vapor escaping during collection: see Needle Selection, Nitrogen Purge, and Collection Compatibility
  • Choosing between a bottle and a cylinder: see Sample Cylinders and Cylinder Service Requirements
  • Valve failing under pressure or temperature at the collection point: see Conditioning Parts and Pressure Management
  • Vapor escaping to atmosphere: see Emission Filters and Needle Selection
  • Installation not fitting the site location footprint: see Enclosures, Pipe Stands, and Station Infrastructure

 

Sample Coolers

The sample usually needs to be below 140 degrees F to meet health, safety, and environment rules for safe handling. Process temperatures over 135 degrees F call for process cooling in most applications. If the cooler does not work, the station cannot be used. Size it with care.

Proper sizing needs process and cooling media properties. These include inlet and outlet temperature and pressure, viscosity, specific heat, thermal conductivity, and specific gravity. Cooling water quality matters too. Scaling and mineral deposits hinder cooling performance.

Bigger is not always better. Over cooling raises viscosity and could lead to plugging.

SENSOR offers two cooler types. The SEN-SC coolers are tube in shell helical style coolers with a tremendous amount of cooling surface area in a compact design. The SENSOR Tube-in-Tube Sample Cooler is a compact stainless steel design for precise cooling with minimal pressure drop. It comes in multiple sizes for a range of viscosities, flow ranges, and temperature conditions.

View the SENSOR Sample Coolers range for sizing guidance.

 

Needle Selection, Nitrogen Purge, and Collection Compatibility

Bending needles, a cap that does not seal, or vapor at the vent could have several likely causes. Check bottle fit in the shroud, cap hole size, septum condition, and the vent connection first.

SENSOR lists needles, septa, and caps under Sampling System Consumable Parts. They still need to match the system. The Sampling System Spare Parts Replacement Guide(PDF) lists the parts and consumables to replace in years 1 through 5 for each system type. Bottle, cap, and septum selection is one of three areas that affect station reliability, along with sample valve design and dead volume. Consistent bottle size and shape ensure good bottle fit and system performance.

Process and vent needle systems for use with a bottle, cap, and septum serve the vast majority of low vapor pressure liquid, chemical, and hydrocarbon streams. The process needle fills the bottle. A separate needle vents it. Replaceable process and vent needles are standard on the SENSOR Basic Bottle Sampling System, which reduces cost and excessive part replacement.

The SENSOR Needle Evacuation System (NES) is the nitrogen purge for bottle stations. It is optional on the Basic Bottle Sampling System and standard on the Pressure Isolating Bottle Sampling System. It includes a check valve, regulator, rotameter, pressure gauge, and block valve. The NES forces out vapor-locked liquid so the sample path has zero dead volume, which could lead to cross-contamination and plugging. For viscous liquids, or liquid trapped in the process needle, the nitrogen purge eliminates the straw effect. It also sweeps vapors through the vent connection, away from the operator and the station. Nitrogen purge is typically set between 3 and 5 psig.

The SENSOR Concentric Needle System aligns the process and venting functions into one single needle. It installs as an option without changing the station design and comes in the same materials as the dual needle system. This needle system works with vials with crimp-on caps and with small bottles under 1 ounce or common lab supply house containers, and can improve station reliability when used in the right applications.

Materials of construction should match what the process stream requires. Standard 316 stainless steel wetted parts suit many streams. Monel 400, Hastelloy C276, Inconel 625, and various other alloys of wetted parts are available. Know the metallurgy, valve packing, and elastomer your stream requires, because most vendors do not specify them for you.

SENSOR punches cap and septum material in-house so it performs well with the needle systems in industrial closed-loop sampling. Laboratory caps have holes too small for industrial needle systems, and their septum material is too thin and easily damaged. Small cap holes can damage one or both needles.

Resealability of septum material is selected to balance damage to the septum against containment of the collected sample. Thin or high durometer septum can lead to poor sealing and exposure issues during transportation to the laboratory.

Bottles should fit snugly, but not tight, in the shroud. A good fit aligns the bottle and cap with the station so the needles pierce the septum properly. A poor fit lets the needles strike the cap and could bend them. Share a sample of your facility’s bottle with SENSOR so fit is confirmed before the station ships. If bottle size, shape, or material requirements change, SENSOR can adapt to meet those changes with a new shroud.

 

Sample Cylinders and Cylinder Service Requirements

Vapor pressure sets the rule. Liquids below 19 psia vapor pressure are typically collected in a bottle. Liquids above 19 psia, and gases and vapors, are typically collected in a cylinder to maintain the sample at process conditions.

SENSOR Sampling provides sample cylinders, pressure-rated containers used to collect, store, and transport gas or high vapor pressure liquid samples from process streams to the laboratory for analysis. They see wide use in oil and gas, petrochemical, environmental monitoring, analytical chemistry, and regulatory compliance applications. Cylinders maintain the integrity of the sample from collection to laboratory analysis by preventing phase changes, contamination, or pressure loss.

  • High-pressure and temperature capability
  • Quick disconnect end connections in a variety of types and brands
  • Safety enhancements such as carry handles and relief valves
  • Various sizes and volumes available
  • Internal outage or dip tubes to ensure proper outage for safe transportation
  • Internal surface treatment and coating to preserve sample integrity
  • Multiple configurations for specialized needs
  • Different materials available to meet specific application needs

Install the station in a fast loop. A fast loop helps keep the sample representative and reduces plugging. Most applications need at least 5 to 8 psig of differential pressure between the high pressure inlet and the lower pressure outlet. Viscous streams may need more differential pressure.

The cylinder system application data sheet(PDF) asks for inlet pressure, fast loop outlet pressure, vapor pressure, viscosity, temperature, particles in the sample, wetted materials, O-ring material, and cylinder size. The standard SENSOR LGSS and VSS systems are designed to operate up to 1440 psig for LGSS and 1800 psig for VSS. High-pressure systems are available.

Wetted parts in cylinder sampling systems are 316 stainless steel as standard. Monel 400 and Hastelloy C276 are available. The sample cylinders themselves are 304 or 304L stainless steel as standard. Options include 316L, Monel 400, Hastelloy C276, Incoloy 825, and Inconel 625. Viton is the standard O-ring material. Viton is not compatible with H2S, and the data sheet recommends Kalrez for H2S service.

Liquids in a cylinder need outage space, which is headspace in the top of the cylinder. A collected sample can increase in temperature during transport, and the liquid then expands into the headspace. Without that space, the cylinder could rupture. The handbook recommends a 20 percent outage, but this can be specified to match any standards that must be maintained. On the LGSS, a sight flow indicator lets the operator obtain proper, safe outage. Dip tubes also set outage, but they are not as reliable due to turbulent flow. Pressure relief valves and rupture discs are last-line safety measures. Gases are compressible, so outage does not apply to gas and vapor systems.

 

Conditioning Parts and Pressure Management

A valve failing under pressure or temperature at the collection point could trace back to a conditioning part that does not match the operating conditions. Conditioning parts manage pressure, flow, and temperature between the process line and the collection point. They are commonly used in conditioning panels for analytical system protection, SWAS systems, and grab sample systems. See the SENSOR Conditioning Parts page for the full lineup.

  • Thermal shut-off valve (SEN-TVS): Keeps process fluids below the maximum allowable temperature to protect analytical instruments and personnel. It is fully mechanical, needs no outside power, and resets automatically when temperature drops. Maximum pressure is 2900 psig and maximum temperature is 356 degrees F.
  • High-pressure reducer valve (SEN-HPRV): Handles inlet pressure up to 5076 psig. Outlet pressure is adjustable from 14.5 to 290 psig.
  • Back pressure relief valve (SEN-BPR): Relieves excess pressure automatically to maintain consistent system performance and protect equipment. It is designed for regulated pressure of 22 psig and a maximum temperature of 167 degrees F (typically downstream of a cooling system).

Match the valve to the actual operating conditions at the sample point.

 

Emission Filters and Cationic Resin Columns

Vapor escaping to the atmosphere, or a cationic conductivity reading that looks wrong, could point to one of two accessories.

SENSOR emission filters capture or remove emissions that would otherwise go to the atmosphere when collecting grab samples. They use highly porous activated carbon to trap chemical pollutants. Color changing indicator spheres, when specified, show when the carbon media is saturated. The BBSS catalogs list the filter canister for use when no vent to flare is available. Connect the vent according to plant guidelines and local environmental standards.

Cationic resin columns support accurate cationic (acid) conductivity measurements in steam and water cycle applications. The columns exchange cations from chemical dosing, including NH4+ from amine treatments and N2H5+ from oxygen scavengers. A built-in color indicator shifts from purple to brown as the resin is consumed. The resin is replaced or regenerated with a 5 percent hydrochloric acid solution.

 

Enclosures, Pipe Stands, and Station Infrastructure

An installation that does not hold up to the site environment, or a panel that is awkward to reach, could trace back to the enclosure, panel layout, or pipe stand specification.

SENSOR enclosures protect sampling components from environmental, safety, and operational hazards. They typically have a hinged door with a latch and drain connections. They mount to a pipe stand or a wall. Housings come in 304 stainless steel as standard, or in 316 stainless steel or glass-fiber reinforced polyester (GRP), with insulation when required. Steam or electric heating is available. In colder climates, a heated enclosure could be needed because cooling water or certain media can freeze or become extremely viscous.

SENSOR pipe stands mount and support instrumentation and panels. Standard material is hot-dip galvanized carbon steel. Options include 316SS, 304SS, and other alloys. Stands use 2-inch schedule 40 pipe, a 10” x 10” x 1/2” thick base plate, two gussets, and a welded cap to keep water out.

Talk to SENSOR about enclosure and pipe stand selection before you order.

 

 

Frequently Asked Questions

Q. What is the difference between the SEN-SC coolers and the SENSOR Tube-in-Tube Sample Cooler?

A. The SEN-SC coolers are tube-in-shell helical-style coolers with a stainless steel double helical coil. The SENSOR Tube-in-Tube Sample Cooler is a compact stainless steel design for precise cooling with minimal pressure drop. Both reduce sample temperature before collection, with a tremendous amount of cooling surface area in a compact design. Overcooling raises viscosity and could lead to plugging. Send SENSOR your process and cooling media properties so the cooler is sized correctly.

 

Q. Does SENSOR supply closed-loop grab sampling system accessories for other manufacturers’ equipment?

A. Yes. SENSOR Sampling offers accessories and spare parts for any sampling system and any manufacturer. Spare parts are available with off-the-shelf speed. The Sampling System Spare Parts Replacement Guide(PDF) lists parts and consumables by system type and year. If you need something not listed on the Parts and Accessories page, contact the factory.

 

Q. What are the SENSOR Cationic Resin Columns used for?

A. They perform cation exchange so operators get an accurate cationic conductivity reading in steam and water cycle applications. Two models are available. The SEN-CR-600 holds 600 cc and the SEN-CR-1600 holds 1600 cc. Flow range is 50 to 250 cc per minute. A color indicator shifts from purple to brown as the resin is consumed.

 

Q. When do I need the SENSOR Concentric Needle System instead of a dual needle?

A. Dual needle systems serve the vast majority of chemical and hydrocarbon streams. Consider the Concentric Needle System if you use vials with crimp-on caps or bottles under 1 ounce, or when using lab supply house containers. It installs as an option on any SENSOR bottle sample system without changing the system design. It can improve system reliability in the right applications. Share your application with SENSOR to confirm the fit.

 

Q. What information does SENSOR need to specify the right closed-loop grab sampling system accessories for my installation?

A. Start with the application data sheet for your station type, bottle or cylinder. It covers pressures, viscosity, temperature, particles, wetted materials, O-ring and valve packing materials, connection details, and container size. When needing a cooler, supply the complete cooler specification data sheet(PDF) with process and cooling water properties.

 

Ready to Specify Your Accessories?

Contact SENSOR Sampling Systems with your process data. SENSOR engineers help match systems and accessories to your service conditions. Spare parts are available with off-the-shelf speed.